9.5.1 Menu→Diagnostic→Sensors
►S1: O2 OK
S2: CO OK
S3: NO OK
Example:
Messages on the state and calibration of the electrochemical sensors:
Ok No problem detected
absent The sensor was not detected
err data Memory data error of the sensor
unknown It is necessary to update the FW of the device
err pos The sensor has been installed in the wrong position
err cal Calibration error (sensor not calibrated)
err curr Currents outside the range
err cfg Do not use this sensor as it has not been accepted on
the screen "types of sensors".
Sensor serial number
Sensor code
Measured Gas
Is sensor current
OK
KEY DESCRIPTION
ESC
Returns to the previous screen.
Selects the available parameters.
OK
Shows the details about the selected sensor.
15/06/18 10:00
►S1: O2 OK
S2: CO OK
S3: NO OK
15/06/18 10:00
S.N. 00000000
Code AACSE15
Gas O2
Range 0-25 %
Is 101.000 uA