EasyManua.ls Logo

E Instruments E500 - End of the Analysis

E Instruments E500
75 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
21
8.1.5 End of the Analysis
-
At the end of the combustion analysis, carefully remove the sample probe and remote air temperature probe, if
used, from their relative ducts, taking care not to get burnt.
-
Switch off the instrument.
Then, proceed to turn off the instrument.
The instrument execute a cleaning cycle, according to what’s set in the menu
“Configuration→Analysis→Autozero→Purging”, during which the pump sucks clean air until reducing the
concentration of CO. The instrument automatically turns off within max. 10 minutes.
Note: It is always advisable to purge the instrument with clean air for at least 5 - 10 minutes before turning it
off.
Gas sampling probe cleaning
-
When you finish using the sample probe clean it thoroughly as described below before returning it to its case:
- Disconnect the sample probe from the instrument and from the water trap (Fig. a-b) then blow a jet of clean air
into the hose of the probe (refer to Fig. b) to remove any residual condensate that may have formed within.
Fig. a
WHEN THE GAS SAMPLING PROBE IS TAKEN OUT OF THE STACK, THE FORMATION OF
SOME CONDENSATION IN THE PROBE TUBE AND IN THE ANTI-
CONDENSATION TRAP
MAY OCCUR.
IT IS ADVISED TO ACCURATELY CLEAN EVERY PART BEFORE PUTTING AWAY THE
PROBE AND THE ANTI-CONDENSATION TRAP IN THE CASE.
IN ORDER NOT TO DAMAGE THE CASE MAKE SURE THAT THE METALLIC PROBE PIPE IS
AT AN INFERIOR TEMPERATURE THAN 60°C.

Table of Contents

Related product manuals