9.2.1 Menu→Measurements→Draft
INTERACTIVE OPERATION FUNCTION
Save
The measure will be printed on the ticket of the current combustion analysis
Print
According to the version of the instrument and/or according with the related
setting, it is possible to print or visualize the QR code
Example:
To measure the draft proceed as follows:
Connect the probe pressure input hose to the instrument P+ input.
Before starting the pressure zeroing sequence pay attention to remove the gas probe from the
stack.
Having carried out the pressure zeroing sequence, insert the probe in the stack and measure the
draft.
OK
Print
Print
15/06/18
10:00
DRAFT
Inlet P+
Draft 0.00 h
Zero sensor
Save Print
KEY
FUNCTION
>
>
Activate the context keys shown on the
display.
ESC
Returns to the previous screen.
OK
Starts the pressure sensor autozero.
15/06/18 10:00
DRAFT
Inlet P+
Draft --- P
Zero
Save
Print
15/06/18 10:00
DRAFT
5
Zero
15/06/18 10:00
DRAFT
Inlet P+
Draft 3.02 P
Zero
Save
Print
15/06/18 10:00
PRINT
Curr. analysis
Copy number 1
Printer BT
QR Code ON
Pairing BT
Print
B 15/06/18 10:00
WARNING
Printing
in progress ...