9.5.2 Menu→Diagnostic→Gas probe
KEY FUNCTION
ESC
Returns to the previous screen.
OK
Starts the test to check the tightness of the gas sampling probe.
Tightness test of the probe.
15/06/18 10:00
GAS PROBE
Calibration
Probe Test
Result: Tight
Results:
Tightness: The system is OK
Leak: Make sure that the probe is connected to the input P- or P+, check the seals of the pneumatic
connections and/or the seal of the condensation trap and check that the test cap is correctly inserted
on the tip of the probe. WARNING: a damaged probe tip may impair the test.
Error: It is not possible to perform the test because the sensor is not calibrated.
Connect the flue gas sampling probe and filter unit
assembly to the instrument;
Fully insert the black rubber cap on the gas probe tip, as
shown in the following picture:
Provided black rubber cap
OK
15/06/18 10:00
GAS PROBE
Close the flue
gas probe
Push OK to
continue
15/06/18 10:00
GAS PROBE
Calibration
15/06/18
10:00
GAS PROBE
Probe Test
15/06/18 10:00
GAS PROBE
Close the flue
gas probe
Push OK to
continue