Calibration procedure
To perform the recalibration the following instruments are needed:
-
Known concentration gas cylinder suitable for the sensor, equipped with a pressure regulator
-
Flow meter
-
Hose with Tee fitting to connect the cylinder to the flowmeter and to the instrument
In the following is described a recalibration example for the CO sensor.
1.
Start the instrument
Keep pressed
for a few
seconds
2.
Once autozero is completed press the button and select the menu Instrument→on site calib.
Insert
WARNING
• Make sure autozero is execute in clean air and terminates correctly.
• Do not connect the gas probe to the instrument.
• Check the battery charge level or connect the power adapter to avoid data loss
during recalibration.
OK
OK
S.n.:0
Fw:1.00
AutoZero in
Air
Air
T: 21.5°C
Keep
►Display
On site calib.
Clock
Bluetooth