The calibration will be possible only when the status is set to ' ---- ' or 'inactive'.
Enter the value of the concentration of the gas applied.
Apply gas to the instrument and adjust the output pressure of the gas from the cylinder so that the flow meter
indicates a minimum flow of 0.5 l/m: this guarantees that the instrument is taking the exact amount of gas required
by the internal pump.
The instrument Measurements the concentration of gas applied; wait at least 3 minutes to allow the reading
to stabilize. The reading is shown in line 'Gas measured'.
Zeroes the timer -
helps to keep under
elapsing during the
stabilization phase.
OK
OK
OK
OK OK
>
>
>
>
control the time
15/06/18 10:00
ON SITE CALIB.
Calibrate
Status ---
El.time
App. gas 100.0 p
Meas.gas 100.0 p
Is 2.22
Ia 0.17
15/06/18 10:00
ON SITE CALIB.
Calibrate
Status active
El.time
App. gas 100.0 p
Meas.gas 100.0 p
Is 2.22
15/06/18 10:00
ON SITE CALIB.
Calibrate
Status inactive
El.time
App. gas 100.0 p
Meas.gas 100.0 p
Is 2.22
15/06/18 10:00
ON SITE CALIB.
Calibrate
Status inactive
El.time 00:03:40
App. gas 1018 p
Meas.gas 0 p
Is 2.22
15/06/18 10:00
ON SITE CALIB.
Calibrate
Status inactive
El.time 00:00:00
App. gas 1018 p
Meas.gas 0 p
Is 2.22
15/06/18 10:00
ON SITE CALIB.
Calibrate
Status inactive
El.time 00:03:00
App. gas 100.0 p
Meas.gas 0 p
Is 2.22
15/06/18 10:00
ON SITE CALIB.
Calibrate
Status inactive
El.time 00:03:50
App. gas 100.0 p
Meas.gas 0 p
Is 2.22
15/06/18 10:00
ON SITE CALIB.
Calibrate
Status inactive
El.time 00:03:50
App. gas 1018 p
Meas.gas 0 p
Is 2.22