The calibration will be possible only when the status is set to ' ---- ' or 'inactive'.
•
Enter the value of the concentration of the gas applied.
•
Apply gas to the instrument and adjust the output pressure of the gas from the cylinder so that the flow meter
indicates a minimum flow of 0.5 l/m: this guarantees that the instrument is taking the exact amount of gas required
by the internal pump.
•
The instrument Measurements the concentration of gas applied; wait at least 3 minutes to allow the reading
to stabilize. The reading is shown in line 'Gas measured'.
Zeroes the timer -
helps to keep under
elapsing during the
stabilization phase.