User manual SEFELEC 5x Series 25 v1.06
Insulation resistance
(SEFELEC 1000-M; SEFELEC 56/506-D/S)
Measurement specifications
Programmable volt per volt from
20 to 1,000 VDC
Generation accuracy
±(1% + 1 V)
(+) pole of the generator
to ground
Dynamic stability for mains ΔV = ±10%: Variation of the
voltage measured
< ±1%.
Maximum current in the measurement circuit:
2 mA - 20% +0%
Maximum capacity of the sample measured
< 100 µF (discharge time < 10 sec.)
Discharge resistance
2.2 kΩ
Resistance measurement range
Measurement scope
(Generator Utest/Umax) x 200
(or 2 TΩ)
Standard version
Voltage
Measurement scope
100 V
100 kΩ to 20
250 V
250 kΩ to 50
500 V
500 kΩ to 100
1,000 V
1 MΩ to 200
Optional 2TΩ version
Voltage
Measurement scope
100 V
100 kΩ to 200
250 V
250 kΩ to 500
500 V
500 kΩ to 1
1,000 V
1 to 2
Generation accuracy
Display with indication of the units, MΩ, GΩ and
2,000 points
Accuracy (in % of the reading, 1U = 1 point of display):
200 GΩ, basic version
±(1.5% + 1U)
Option - 2 TΩ and U <= 200 Vdc
±(2% + 1U)
Option - 2 TΩ and U > 200 Vdc
±(1% x U test / 100 + 1U)
Capacity mode
1.00 MΩ at 200
Accuracy
Precision mode NORMAL ± 100
Input impedance
10 MΩ ±1%