XS series operating manual
XS user manual v0.55 - 92 -
12 Input output Interfaces
Input-Output for Programmable Logic Controller Interface (PLC)
This option gives a functional control of the measurement unit by a Programmable Logic Controller
system.
WARNING
The PLC interface must be selected in the SETUP menu, line INTERFACE :PLC
Electrical specifications
INPUTS :
Number : 7
Type : Optoelectronic
Input resistance : 1.5 kohms
Minimum voltage : 11 VDC or 8 VAC
Maximum voltage : 43 VDC or 30 VAC
OUTPUTS:
Number : 5
Type : Dry contacts
Maximum voltage : 70 VDC
Switching rating : 30W
Maximum current : 0.15 ADC
Description of the logical states
Input :
Logical state HIGH : DC or AC voltage with an amplitude between Umin and Umax.
Logical state LOW : no voltage.
Output :
Logical state HIGH : closed contact.
Logical state LOW : opened contact.
.