VACUUM SETTING
From the SETUP, select VACUUM SETTINGS, default setting is displayed:
Allows to modify the default vacuum time and the default time of check.
Each value can be modified, within the values shown in parentheses.
NOTE: press DEFAULT to restore default values:
Vacuum time 25 min
Time of check 2 min
Vacuum rising 0,1 mbar
OPTIONS
Select OPTIONS, insert the OPTION code:43210791 then press ANALYZER, the following
screen will be displayed:
Choose ON to enable gas analyzer or OFF to disable it then press ENTER.
VACUUM SETTINGS
100
2
25
DEFAULT
SETUP
DD/MM/YY XX:XX PM
ENTER
ESC
Vacuum time
(Min 2, Max 120 min)
Vacuum rising
(min 100, max 900 mbar
Time of check
(Min 1, Max 120 min)
DD/MM/YY XX:XX PM
ENTER
ON
OFF
ENABLE ANALYZER