H100N Pulse Oximeter User Manu Specification
- 103 -
A1.3 Parameters
Measurement range
SpO
2
1 % ~ 100 %
PR 20 bpm ~ 300 bpm
Perfusion range 0.03 % ~ 20 %
TEMP 0 ~ 50
Accuracy Tolerance
Saturation
Adult
70 ~ 100 ± 2digits
Neonate
70 ~ 100 ± 3digits
Low Perfusion
70 ~ 100 ± 2digits
Pulse Rate
Adult and Neonate 20 bpm ~ 300 bpm ± 3digits
Low Perfusion 20 bpm ~ 300 bpm ± 3digits
Data Update Period 7 seconds
TEMP 25 ~ 45 ± 0.1℃℃