EDAN H100N - A2.2 Electromagnetic Immunity - for All EQUIPMENT and SYSTEMS

EDAN H100N
130 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
H100N Pulse Oximeter User Manu EMC Information
- 107 -
/flicker emissions
IEC/EN61000-3-3
supplies buildings used
for domestic purpose.
A2.2 Electromagnetic Immunity - for all
EQUIPMENT and SYSTEMS
Guidance and manufacturer’s declaration – electromagnetic
immunity
The oximeter is intended for use in the electromagnetic
environment specified below. The customer or the user of the
oximeter should assure that it is used in such an environment.
Emissions test Compliance
Compliance
level
Electromagnetic
environment -
guidance
Electrostatic
discharge(ESD)
IEC/EN61000-4-2
±6kV contact
±8kV air
±6kV
contact
±8kV air
Floors should be
wood, concrete
or ceramic tile. If
floors are