H100N Pulse Oximeter User Manual Oximeter Operation
- 62 -
Function button to enter setup. The Temp Hi Alarm
box changes from real line to broken line. The adjustable
range of the upper limit of Temp Alarm is from “0.1 +
the lower limit of Temp Alarm” to 50.0.
Press the Up Arrow or the Down Arrow button to
increase or decrease values.
Choose Temp Lo Alarm in menu, and press the
Function button to enter setup. The Temp Lo Alarm
box changes from real line to broken line. The adjustable
range for the lower limit of Temp Alarm is from 0 to
“the upper limit of Temp Alarm - 0.1℃”.
Press the Function button, and confirm the alarm
range setup.
Hi Alarm is always higher than Lo Alarm by at
least 1.
Press the On/Off button to exit the menu, and
return to measurement state.