H100N Pulse Oximeter User Manual Oximeter Operation
- 59 -
on again.
Press the Up Arrow or Down Arrow button to
increase or decrease values.
Choose SpO
2
Lo Alarm in the menu, press the
Function button to set it. The SpO
2
Lo Alarm box will
change from real line box to broken line box. The
adjustable range for the lower limit of SpO
2
Alarm is
from 0 to “the upper limit of SpO2 Alarm - 1”. If the
value of SpO2 Lo Alarm is set to less than 85, it will
restore to 85 after the oximeter is turned on again.
Press the Up Arrow or Down Arrow button to
increase or decrease values.
SpO
2
Hi Alarm is always higher than SpO
2
Lo
Alarm by at least 1%.
Press the Function button, and confirm the alarm
range setup.
Press the On/Off button to exit the menu, and
return to measurement state.