H100N Pulse Oximeter User Manual Oximeter Operation
- 58 -
3.3.10 SpO2 Alarm Setup
The user can choose SpO
2
Hi Alarm and SpO
2
Lo
Alarm in menu to adjust SpO
2
alarm limit. Press the Up
Arrow button or Down Arrow button to increase or
decrease alarm limit.
By default, SpO
2
Hi Alarm and SpO
2
Lo Alarm in Neo
mode are set to 95 and 90 respectively; while they are
100 and 90 in Adu mode respectively.
Set the SpO
2
alarm limits as follows:
Choose SpO
2
Hi Alarm in the menu, and press the
Function button to enter setup. The SpO
2
Hi Alarm box
will change from real line box to broken line box. The
adjustable range for upper limit of SpO
2
is from “1 + the
lower limit of SpO
2
” to 100. If the value of SpO2 Hi
Alarm is set to less than 85, it will restore to default
value after the oximeter is turned on again. In the NEO
mode, if the value of SpO2 Hi Alarm is set to higher
than 95, it will restore to 95 after the oximeter is turned