H100N Pulse Oximeter User Manual Introduction
- 35 -
Checking the Ni-MH Battery Package
The performance of a rechargeable Ni-MH battery
package may deteriorate. To check the performance of
the battery, follow the procedures below:
1. Disconnect the pulse oximeter from the patient and
stop all monitoring and measuring procedures.
2. Place the pulse oximeter in the charger stand and
connect the AC mains. Allow the battery to be charged
uninterruptedly for above 2.5 hours.
3. Disconnect AC mains and allow the pulse oximeter
to run in the measurement state until it shuts off.
The operating time of a battery reflects its performance
directly. If the operating time of a rechargeable Ni-MH
battery package is noticeably shorter than that stated in
the specifications, replace it or contact your service
personnel.