SE-1 Electrocardiograph User Manual Safety Guidance
- 7 -
CAUTION
j) Test the patient leakage current under single fault condition with mains voltage on the
applied part according to IEC/EN 60601-1: Limit: 50μA (CF)
The data should be recorded in an equipment log. If the device is not functioning properly
or fails any of the above tests, the device has to be repaired.
6. Ruptured fuse must only be replaced with that of the same type and rating as the
original.
1.2.4 Cleaning & Disinfection Cautions
CAUTION
1. Turn off the power before cleaning and disinfection. If the mains supply is used, the
power cord should be dragged out of the outlet. Prevent the detergent from seeping
into the equipment.
2. Do not immerse the unit or the patient cable into liquid under any circumstances.
3. Do not clean the unit and accessories with abrasive fabric and avoid scratching the
electrodes.
4. Any remainder of detergent should be removed from the unit and the patient cable
after cleaning.
5. Do not use chloric disinfectant such as chloride, sodium hypochlorite etc.
1.3 List of Symbols
No. Symbol Description
1
Output
2
Input