SE-1 Electrocardiograph User Manual Settings
- 36 -
Setting Method:
1. Press the ID key (the up or down arrow) to move the arrow
on the right of the LCD screen to the item to be changed.
2. Then press the M/F or AGE key to select an option for the item.
3. After setting the items, press the MENU key to exit the system setup interface.
Note: If you set DEFAULT SETTING to RESTORE, all the items will be restored to the
default settings except the date, time and language.
The following sections provide an introduction to the menu items.
8.1 Filter Settings
You can set FILTER SETTING to EMG, AC, ALL (both EMG and AC) or OFF (no filter).
When it is set to OFF, the filters will not work. Generally, it is recommended to be set to ALL
to get a better ECG record.
Note: To pass the distortion test, the electrocardiograph has to be configured with the
highest bandwidth in filter settings. Otherwise, ECG signal may be distorted.
8.2 Recording Settings
PWAVE START : ON
RECORD LENGTH : 3
RECORD COUNT : SECOND
: ALL
: ON
: 3
: SECOND
FILTER SETTING
PWAVE START
RECORD LENGTH
RECORD COUNT