SE-1 Electrocardiograph User Manual Settings
- 37 -
RECORD SPEED : 25
HEARTRATE PRINT : ON
Take the above settings for example, the ECG will be printed from a P wave, the printing speed
is 25mm/s, the printing length of each lead is 3 seconds, and the heart rate will be printed out
on the bottom of the recorder paper. If RECORD COUNT is set to QRS, the printing length of
each lead is three QRS waves.
Note: The printing duration of each lead must be more than 2 seconds. When
RECORD COUNT is set to QRS, if the time period of QRS waves is so short
that the printing duration of each lead is less than 2 seconds, printing will
continue until it reaches 2 seconds.
8.3 Date and Time Settings
YEAR : 4
MONTH : 8
DAY : 6
HOUR : 14
MINUTE : 25
As the above settings show, the date and the time is set to Aug. 6
th
, 2004, 14:25 PM. The time
will be displayed on the LCD screen. The date and the time will be printed in the ECG report.
Note: Please set DATE&TIME correctly when it's the first time you use the
electrocardiograph.
8.4 Print Head Test
PRINT HEAD TEST : OFF
The print head test is used to check whether the print head can work normally. By default,
PRINT HEAD TEST is set to OFF. After the recorder paper is loaded, set PRINT HEAD
TEST to ON by pressing the M/F or Age key. Then the triangle waves in effective paper
width will be printed. The status of the print head can be estimated from these triangle waves.
Press the M/F or Age key to stop the printing test.