© Edwards Limited 2011. All rights reserved. Page 7
Edwards and the Edwards logo are trademarks of Edwards Limited.
Technical data
D395-93-880 Issue A
2.4 Electrical data
If your T-Station uses an E2M1.5, the motor start-up current is drawn for less than one second, so you must use slow-
blow fuses to prevent unnecessary fuse failure when the pump starts. If you use the pump at temperatures lower
than 12 °C (53.6 °F), the start-up current will be drawn for longer; this may cause the motor thermal overload device
to open.
Table 2 - Active gauge connector pin-out
Pin Function
1 Power supply positive
2 Power supply common
3Signal input
4 Identification
5Signal common
6 Control line 1
7 Control line 2
8N/C
Table 3 - Electrical data
Pump combination
Supply Voltage
50/60 Hz
Current (A)
Power (W)
Full load Start
XDD1 / EXT75DX
100 3.1 -
310
120 2.6 -
200 1.6 -
230 1.35 -
E2M1.5 / EXT75DX
100 4.7 13.6
470
120 3.92 12.0
200 2.2 6.4
430
230 1.9 5.7