Operation
4.3.2 Gauge screen
When the Gauge screen is selected, the Gauge LED is lit and the gauge pressure reading is displayed in the selected
units. If no gauge is connected the display shows "---".
4.3.2.1 Connecting a gauge
When a gauge is first connected the display shows "ID" followed by a number to identify the new gauge. If the gauge
is an ASG, the display then goes to ASG Range Select, otherwise the display reverts to showing the pressure reading.
If the gauge type is not supported the display shows "???". When no gauge is connected the display shows "---".
4.3.2.2 ASG range select
When an ASG is connected the display changes to select the ASG range. 1000 mbar is assumed by default, and the
display flashes "1.0
3
". Press the NEXT key to select between 1000 and 2000 mbar, then press the ENTER key to
confirmed the selection.
4.3.2.3 Gauge ON/OFF control
Gauges which support ON/OFF control (e.g. AIM gauges) can be turned ON and OFF using the ENTER key. When first
connected the gauge is in the OFF state and the display shows "OFF". When the gauge is turned on the display will
show "Str" whilst the gauge is starting up, and will then display pressure.
4.3.2.4 Gauge screen key actions
Keys Short Press Long Press
START Go to Turbo Screen -
ENTER ON/OFF control of supported gauge or
Acknowledge Error
-
NEXT Go to Vent Control Screen Scroll through menu screens
Turbo
Gauge
Gauge
On/Off