98
2. When the load is connected, press TEST to start the calibration process.
3. Enter Standard Current Reading.
Calibration of 350nA DC Current Range
1. Connect the 1G load in series with the standard current meter.
2. When the load is connected, press TEST to start the calibration process.
3. Enter Standard Current Reading.
Calibration of 8V AC Voltage Range
1. Connect the standard 8.00VAC meter from Current to Return terminal.
2. When the leads are shorted, press TEST to start the calibration process.
3. Enter Standard Voltage Reading.
Calibration of 30A AC Current Range
Step1
1. Connect the standard current meter and series 100mΩ load from Current to Return.
2. When the standard current meter is connected, press TEST to start the calibration process.
3. Enter Standard Current Reading.
Step2
1. Connect the Current and Return terminal.
2. When the leads are shorted, press TEST to start the calibration process.
3. Automatic calibration processing.
Calibration of GFI 5.0mA
Step1
1. Connect the standard 200k with the standard AC current meter from H.V. to Case.
2. When the standard load is connected, press TEST to start the calibration process.
3. Enter Standard Current Reading.
Step2
1. Connect the standard 200k with the standard DC current meter from H.V. to Case.
2. When the standard load is connected, press TEST to start the calibration process.
3. Enter Standard Current Reading.
Calibration of ACW ARC AC
Connect to ARC fixture point 19 and press TEST to start the calibration process.
Calibration of DCW ARC DC
Connect to ARC fixture point 12 and press TEST to start the calibration process.
Calibration Date 11/05/2014
Enter the date of Calibration.