58
4.7.1 ACW
Touch ACW for the test type and use the enter key (↵) to save the Test Type and move to the next
test parameter:
Note: The Offset value can only be set by performing the auto offset. To perform an auto offset
disconnect the DUT and leave all the test leads open. Press the test button on the front panel.
Reading will be displayed on the screen when the instrument is performing the auto offset. Once
the instrument is has finished performing the auto offset the numeric value will be displayed on
the screen and will be automatically saved to be used when the actual test is performed.