68
Selecting Delete All function will simply delete all the test results from the internal memory.
Selecting Delete will delete a selected step.
6.1 Error Messages and Fail Messages
If there is an error during testing, the error description will be displayed on the screen. Below is a
list of error messages that the SE reports:
Abort: This message appears on the display if the test in process is aborted with the RESET button
or remote Reset control.
Hi-Limit: This message appears on the display if the DUT measurement exceeds the Hi-Limit
setting of any parameter.
Lo-Limit: This message appears on the display if the DUT measurement drops below the Lo-Limit.
CONT-Fail: This message appears on the display if the DUT fails the basic continuity check
performed during an AC/DC Withstand test (if Continuity is selected “ON”).
Arc-Fail: This message appears on the display if the DUT arcing current exceeds the Arc Sense limit
and Arc function is active (Arc Sense = 1…9) of the AC/DC Withstand test.
Short: This message appears on the display if the DUT current is well beyond the metering range
of the test.
Charge-LO: This message appears on the display if the leakage current during Ramp-up falls below
the Charge -LO setting.
Breakdown: This message appears on the display if the DUT current is well beyond the metering
range of the test and the arcing condition beyond the arc sense limit.
GND-Fault: This message appears on the display if the GFI threshold is exceeded during the test.