5. Test Connections
5.1 Connecting the Test Leads
The instrument comes with all cables necessary for performing a Hipot and Continuity test. Plug
the red alligator clip into the HV receptacle on the SE. Connect one of two black alligator clip leads
to the Return receptacle and the other to the Cont. Check receptacle.
5.2 Connecting the Adaptor Box
The adapter box provides an easy way to connect a line cord-terminated DUT to the SE.
5.3 Interlock Connector
SE is equipped with a Remote Interlock feature. Remote Interlock utilizes a set of closed contacts
to enable the instrument’s output. If the Remote Interlock contacts are open the output of the
instrument will be disabled. Remote Interlock can also be referred to as a remote system lockout,
utilizing “fail when open” logic. To disable the Remote Interlock feature connect the Interlock Key
into the Signal Input port located on the back of the tester.