83
SP <prompt message>
Adds or edits a prompt message for the active step.
SP
Removes or deletes the prompt that had been created for the active step.
8.4.4 Test Parameter Editing Commands and Companion Queries
These commands are used to modify the test parameter within each step. These commands
require a parameter value to be included with the command. The companion query command will
read the parameter. The writing of the parameter requires that the unit not be included with the
value, only the numeric value should be included with the command. Also, when the query
commands are used the response will not include the unit’s characters. Many of the commands
will function the same way for multiple test types; however, the input range may be different and
therefore used a different possible set of values.
COMMAND
NAME
TEST
TYPES
VALUE
ED {1|0}
ED?
Edit Current Meter
Display
ACW
1 = Real, 0 = Total
EV <value>
EV?
Edit Voltage
ACW
DCW
IR
GND
1 - 5000V
1 - 6000V
1 - 6000V
3.00 - 8.00V
EC <value>
EC?
Edit Current
GND
1.00 - 32.00A
EPO n
EPO?
Edit Current Polarity
DCW
IR
0 = +
1 = -
ECG < value >
ECG?
Edit Charge-Lo
DCW
IR
0.0 - 350.0uA
0.000 - 3.500uA
ECT {1|0}
ECT?
Edit Continuity
ACW
DCW
1= On, 0=Off
ERU < value>
ERU?
Edit Ramp-Up
ACW
DCW
IR
0.1 - 999.9s
ERD <value >
ERD?
Edit Ramp-Down
ACW
DCW
IR
0.0 - 999.9s
0, 1.0 - 999.9s
0, 1.0 - 999.9s
ERH n
ERH?
Edit Ramp-Hi n
DCW
n=0-10000uA