11 EMC Tables
27 / 31
The Fast-Pack pro is intended for use in the electromagnetic environment specified
below. The customer or the user of the
Fast-Pack pro should assure that it is used
Immunity
test
IEC 60601 test
level
Compli
ance
level
Electromagnetic
environment - guidance
Conducted dis-
turbances
induced by RF
fields
IEC 61000-4-6
Radiated RF EM
fields
IEC 61000-4-3
.
Proximity fields
from RF wireless
communication
equipment
3 V
0.15 MHz – 80 MHz,
6 V in ISM bands be-
tween 0.15 MHz and
80 MHz, 80 % AM at
1 kHz
3 V/m, 80 MHz – 2,7
GHz, 80 % AM at 1
kHz
See the RF wireless
communication
equipment table in
"Recommended
minimum separation
3 V
3V/m
Complies
Portable and mobile RF
communications equipment should
be usedno closer to any part of the
Fast-Pack pro, including cables, than
the recommended separation
distance calculated from the equation
applicable to the frequency of the
transmitter.
Recommended minimum
separation distances
See the RF wireless communication
equipment table in "Recommended
minimum separation distances"
Recommended minimum separation distances
Nowadays, many RF wireless equipments have being used in various healthcare
locations where medical equipment and/or systems are used. When they are used in
close proximity to medical equipment and/or systems, the medical equipment and/or
systems’ basic safety and essential performance may be affected. The
Fast-Pack pro
has been tested with the immunity test level in the below table and meet the related