Maintenance
User Manual Hemo Control
8. Maintenance
8.1 Cleaning and disinfection
8.1.1 Mains adaptor, housing and display
Cleaning the mains adaptor, housing and display is best accomplished
with a cloth, lightly dampened with clear water. For more stubborn
soiling, a mild soap solution may be used. For disinfection, commercially
available solutions intended for surface disinfection can be used if they
do not contain alcohol or other solvents. EKF-diagnostic GmbH
recommends the use of Kohrsolin® FF. Please note the application
instructions of the Manufacturer.
! Note !
To avoid operation of buttons when cleaning the touchscreen,
move the cuvette holder into half open position.
8.1.2 Cuvette holder
The cuvette holder can be removed from the device for cleaning.
Proceed as follows:
Open the cuvette holder until you feel
resistance.
Press down the latch on the left-hand
side of the cuvettte holder with a ball-
point pen and concurrently retract the
cuvette holder to the front side.
Figure 24 - Cuvette holder
The cuvette holder can be cleaned with a mild soap solution. For
disinfection commercially available, solvent-free agents intended for
surface disinfection can be used.
Wait until the cuvette holder is completely dry before reinserting it into
the device. For reinserting the cuvette holder, push it in the correct
position into the opening in the housing until it has noticeably engaged.