Technical Data
User Manual Hemo Control
31
10. Technical Data
10.1 Hemo Control Photometer
Measuring procedure
Optical absorption photometry
Source
Dual-color LED 570 / 880 nm
Dominant wavelengths
1st Wavelength: 570 5 nm
2nd Wavelength: 880 10 nm
Spectral half-value widths
1st Wavelength: 15 3 nm
2nd Wavelength: 50 nm
Measuring range
0 256 g/L
Linearity
Measured value 0 200 g/L: 3 g/L
Measured value > 200 g/L: 7 g/L
Variation coefficient
2%
Correlation coefficient
with reference to NCCLS
method
0.98
Sample material
Venous, arterial or capillary human
blood
Sample carrier
Hemoglobin Microcuvette
Mean measuring time
25 60 s depending on concentration
Interfaces
RS232
Measured value memory
4000 measured values
Ambient temperature
Room temperature 15 40 °C
Transport temperature
-20 to +50°C
Humidity
10 85 % relative humidity