13
8
7
9
2.4 Checking the ambient temperature
To ensure that equally reliable test results are
achieved under dierent climatic conditions,
the measuring device is equipped with a
temperature sensor. The temperature sensor
is located underneath the opening for the
sensors
9
.
2.3 Checking the code setting
The code is displayed briefly when you
switch measuring mode on
7
. If more than
25 tests are performed with the same code, a
reminder to check the code will be displayed
for three seconds
8
. Pressing the OK button
within three seconds will take you to the code
configuration menu.
WARNING
Make sure you do not touch or cover the
temperature sensor with your hand. Failure to
do so results in the risk that the temperature
sensor will read the ambient temperature
incorrectly and the measuring device may
record incorrect test results.