ATTENTION
Never bring the opening of the test
solution bottle directly into contact
withthe sensor. Failure to do so will
contaminate thecontents of the bottle
and render it unusable.
Test solution, sensor and measuring device
must be at the same temperature when
performing the functionality test.
WARNING
Do not swallow the test solution.
Avoid contact with mucous membranes.
Risk of small parts being ingested:
keep the test solution out of the reach
of children.
3.3 Functionality test
The functionality test is used to check the
correct functioning of the Lactate Scout 4
measuring system. If there are any doubts
regarding the accuracy of the test result or
the correct functioning of the measuring
device, you must perform a functionality test.
Please use the Lactate Scout test solution for
the functionality test. The label on the test
solution bottle states the range of the test
solution.
The test solution is available in the following
concentrations:
8.9 - 1 1.1 mmol/L (display: 10 mmol/L)
4.5 – 5.6 mmol/L (display: 5 mmol/L)
1.8 - 2.2 mmol/L (display: 2 mmol/L)
The test solution can be used for a period
of three months after opening for the first
time. Once opened, ensure thetest solution is
stored tightly closed between 15 – 25°C.