Message
Cause / Action
#1 - Probe Probe to gauge communication failure.
Integral Gauges: Contact Elcometer.
Separate Gauges: Remove probe and refit. If error
persists, contact Elcometer.
#2 - Probe
Corrupt data output from probe.
Integral Gauges: Contact Elcometer.
Separate Gauges: Replace probe. If error persists,
contact Elcometer.
#3 - Probe Internal Error - contact Elcometer.
Probe Unsupported
No probe connected. Connect probe.
Probe is Dual FNF but gauge is ferrous or non-ferrous
only or Mk 3 probe is being used with a Mk 4 gauge.
Use correct probe.
Probe Changed
Probe has been changed during a calibration
procedures. Calibration adjustment required.
Unstable Reading
External electromagnetic interference or gauge is faulty.
Isolate from electromagnetic interference. If error
persists, contact Elcometer.
Value Too Large
Calculated value is too large to be displayed. Contact
Elcometer.
Calibration Error Incorrect calibration calculation. Re-calibrate. If error
persists, contact Elcometer.
Bluetooth PC Reply
Bluetooth link working at limit of its range. PC is busy
running other applications. Interference. PC
application doesn’t send replies. Move gauge closer to
PC. Shutdown other applications. Remove source of
interference. If error persists, contact Elcometer.
Error - EEPROM Write
Gauge EEPROM communication error.
If problem persists, EEPROM is faulty.
Error Codes
Under certain conditions, the gauge will display error messages. These
message are cleared by pressing any key. The cause of the error should
be corrected before proceeding.