SECTION 1
INTRODUCTION
1.1 Cytopro Cytocentrifuge Overview
Table 1: Cytopro Specifications
Table 2: Cytopro Rotor Specifications
Single: 0.5 mL max
Dual: 2 x 0.3 mL
Cytopro Magnum: 6 mL
Single = 38.5 mm
2
(7 mm diameter)
Dual = 77 mm
2
(2 x 7 mm diameter)
Cytopro Magnum = 315 mm
2
Up to 8 slides and Cytopro chambers
Rotor Dimensions (Diameter x Height, including lid)
22.6 x 6.2 cm (8.9 x 2.4 inches)
Rotor Weight (including lid)
*Do not overfill cytocentrifuge chambers. See Section 3.1 or the Cytopro Methods Manual for detailed
instructions and warnings.
100 to 240 VAC @ 50 to 60 Hz
2.0 Amp time-lag (Type T) (5 x 20 mm) (2 required).
Manufactured by Littlefuse (Part No. 218002) or
Bussman (part No. GDC-2A)
Ambient Temperature
Operating
Storage
Lid Interlock: Lid must be closed to begin a cycle, and
remains locked down while rotor is spinning
7 in. WVGA (800 x 480 pixels) TFT
Dimensions
Width
Height (lid closed)
Depth
Height (lid open)
57 cm (22 in.)
25 cm (10 in.)
54 cm (21 in.)
58 cm (23 in.)