Trouble shooting / List of malfunctions
12
SM/Transsonic_LC/1104/D © Elma GmbH & Co KG
Malfunction 1. Check:
Heating
switch
- if lighted
2. Check:
Fuse
printed
board
Possible causes Remedy
Fuse automatic
(building)
released;
No heating;
No ultrasound
OFF faulty short circuit in
heating
disconnect heating
cable,
check unit again (new
fuses)
if short circuit does not
occur again, the
heating is faulty:
replace faulty heating
No heating
(Ultrasound OK)
OFF OK faulty heating switch
replace component
No heating
(Ultrasound OK)
OK OK faulty thermostat
button (check with
ohmmeter)
replace component