Trouble shooting / List of malfunctions
© Elma GmbH & Co KG SM/Transsonic_LC/1104/D
17
6.4 T780 – T1060H
Malfunction 1. Check:
Pilot
lamp
ultra-
sound
2. Check:
Heating
switch
- if
lighted
3. Check:
Fuse
inter-
ference
filter
4. Check:
Fuse
printed
board
Possible
causes
Remedy
Check the unit for possible damages and traces of liquid inside (e.g. leak in the
tank, liquid/humidity entering from outside)
No ultrasound
/ weak
ultrasound
(Heating OK)
OK OK OK OK
secondary
transformer
coil
output
capacitor
scorching in
transducer
system
replace
component
replace
compl.
printed
board
open/
check
transd.
system
No ultrasound
(Heating OK)
OK OK OK faulty
printed
board:
(transistors
diodes)
replace
component
replace
compl.
printed
board
No ultrasound
(Heating OK)
OFF OK OK OK
interference
filter
timer
replace
component
replace
compl.
interf. filter
No ultrasound
No heating
OFF OFF OK OK
interference
filter
mains cable
replace
component
replace
compl.
interf. filter
replace
mains
cable