EM TEST VDS 200 Series
Manual for Operation V 5.19 21 / 57
3.3.2.3.
Supply voltage profile
This test is to determine the reset behavior of the device under Test at different voltage drops. This test is
applicable to equipment with reset function.
Parameters:
Vb
0.0V
-
30.0V (60.0V)
Va1
- 30.0V (- 60.0V)
-
30.0V (60.0V)
Va2
- 29.9V (- 59.9V)
-
30.0V (60.0V)
Ve
- 30.0V (- 60.0V)
-
30.0V (60.0V)
dVa
- 30.0V (- 60.0V)
-
30.0V (60.0V)
t1
0.1s
-
99.9s
t2
0.1s
-
99.9s
t3
0.1s
-
99.9s
n
1
-
30,000 / endl.
tri
Auto / Manual
I
1A
-
30A (Imax)
3.3.2.4.
Pulse ’Starting profile’
This test is simulate a motor startup including a possible ripple.
tr, tf = 10ms
Ripple = 2Hz
Limits depends VDS voltage range
Input restrictions
0.0 V Vb + Va1 30.0V (60.0V)
0.0 V Vb + Va2 30.0V (60.0V)
Parameters:
Vb
0.0V
-
30.0V (60.0V)
Va1
- 30.0V (- 60.0V)
-
30.0V (60.0V)
Va2
- 30.0V (- 60.0V)
-
30.0V (60.0V)
t1
0.1s
-
999ms
t7
5ms
-
999ms
t8
5ms
-
999ms
t9
0.5s
-
99.5s
t11
5ms
-
999ms
n
1
-
30,000 / endl.
tri
Auto / Manual
I
1A
-
30A (Imax)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EM TEST VDS 200 Series and is the answer not in the manual?

EM TEST VDS 200 Series Specifications

General IconGeneral
BrandEM TEST
ModelVDS 200 Series
CategoryAmplifier
LanguageEnglish

Related product manuals