EasyManua.ls Logo

Emerson Bristol 2808 Series - Elevated Zero Calibration; Zero-Based Calibration

Emerson Bristol 2808 Series
140 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3-8 / Calibration 2808B
Suppressed Zero Calibration:
0 psi < 1V (4-20mA) output [0 psi results in an output less than 1V (4mA)]
Sample Ranges:
1 to 10 psi
10 to 60 psi
50 to 100 psi
The above procedures are described in topics 3.6, 3.7, and 3.8. Select the procedure you
require.
3.6 ZERO-BASED CALIBRATION
This procedure describes the process by which the zero and span settings are obtained for
the 1-5V and 4-20mA output stages of the 2808 DP and PT transmitters.
The following parameters relate directly to the calibration process:
URL: Upper Range Limit of sensing element. The URL is the maximum input
pressure that can be applied without over-pressuring the sensing
element.
Span: The algebraic difference between the limits of the range (P
max
– P
min
).
The desired span must always be less than or equal to the upper range
limit (URL) of the sensing element. Span is expressed as a percentage of
the upper range limit of the sensing element.
Zero: The point within the available pressure range the user defines as a zero
pressure reference.
P
max
: Maximum input pressure of a desired range, not necessarily the
maximum range of the sensing unit.
P
min
: Minimum input pressure of a desired range.
Theory of Operation: The output stage can potentially take a user defined portion of the 0-
100% URL pressure input, and display it as a corresponding 1-5V or 4-20mA output. The
range, desired by the user, is obtained by setting the appropriate coarse and fine zero and
span settings. Each coarse setting is described in Tables 3-A and 3-B, whereas the fine
settings are adjusted manually until the desired output is achieved. Fine adjustments are
made with two external adjustment screws. The output stage is designed to yield a Turn
Down of better than 6, while the fine adjustment screws should affect the output zero and
span by no more than 1.5%/Turn.
3.7 ELEVATED ZERO CALIBRATION
1. Calculate the total span required for the elevated zero range. For example, if the desired
elevated zero range is -10 to +30 psi, the total span will be 30-(-10) = 40 psi.

Table of Contents

Related product manuals