4 Installation requirements and
limitations
4.1 Requirements and limitations for installation
Follow the steps below to access the wiring connectors.
1.
WARNING!
ELECTROCUTION HAZARD
Disconnect from supply before opening enclosure. Keep cover tight while circuits are
alive. Conduit seals must be installed within 450mm (18") of the enclosure.
Failure to follow these instructions may result in death or serious injury.
Open the enclosure to access the wiring connectors. The jam screw on the cover
may need to be loosened. Remove the internal electronic assembly from the
enclosure to access the electrical connectors.
2. Remove the two captive screws. Disconnect the ribbon cable from the display
module and set the display module aside.
3. Disconnect the wires/cables from the electronic board.
4. Install in accordance with applicable local and national regulations (e.g. NEC).
5. Install and service should be performed only by trained service personnel.
6. Service requiring replacement of internal components (not including battery, if
equipped) must be performed at the factory.
7. Use suitably certified and dimensioned cable entry device and/or plug.
8. All input circuits must be derived from a CSA Approved Class 2 source.
Installation requirements and limitations
User manual 25