Table 2-2: Maintenance (continued)
Daniel MeterLink utility Diagnostics Action(s)
transducer face, or path
length configured
incorrectly).
Adjust SSMin or SSMax only
if other checks pass (consult
a Daniel Service
representative before
making these adjustments).
Meter Monitor (Detailed) view
Meter Data List
Electronics Temperature of
out range
Temperature of the
electronics is out of
nominal operating range
below -40 °C or above 100
°C (-40 °F or above 212 °F).
Heat or cool the meter
electronics housing. If
operating temperature
exceeds 60oC, remote
mount the Transmitter
Electronics Enclosure.
Meter Monitor (Detailed) view
Meter Data List
Electronics voltage out of
range
Systems voltages are valid if
1.0V, 1.2V, 2.5V, 3.3V or
the Acquisition Module
valid voltages are 1.2V,
2.5V or 3.3V.
Replace the CPU Module if
one or more of the System
Voltages is out of range.
Replace the Acquisition
Module if one or more of
the voltages is out of range.
Troubleshooting Maintenance and Troubleshooting manual
June 2019 P/N 3-9000-769
40 Gas Ultrasonic Flow Meters

Other manuals for Emerson Daniel 3410 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson Daniel 3410 Series and is the answer not in the manual?

Emerson Daniel 3410 Series Specifications

General IconGeneral
BrandEmerson
ModelDaniel 3410 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals