Table 2-2: Maintenance (continued)
Daniel MeterLink utility Diagnostics Action(s)
transducer face, or path
length configured
incorrectly).
• Adjust SSMin or SSMax only
if other checks pass (consult
a Daniel Service
representative before
making these adjustments).
Meter Monitor (Detailed) view
Meter Data List
• Electronics Temperature of
out range
• Temperature of the
electronics is out of
nominal operating range
below -40 °C or above 100
°C (-40 °F or above 212 °F).
— Heat or cool the meter
electronics housing. If
operating temperature
exceeds 60oC, remote
mount the Transmitter
Electronics Enclosure.
Meter Monitor (Detailed) view
Meter Data List
• Electronics voltage out of
range
• Systems voltages are valid if
1.0V, 1.2V, 2.5V, 3.3V or
the Acquisition Module
valid voltages are 1.2V,
2.5V or 3.3V.
• Replace the CPU Module if
one or more of the System
Voltages is out of range.
• Replace the Acquisition
Module if one or more of
the voltages is out of range.
Troubleshooting Maintenance and Troubleshooting manual
June 2019 P/N 3-9000-769
40 Gas Ultrasonic Flow Meters