WARNING!
EXPLOSION HAZARD
Substitution of components may impair the intrinsic safety and cause ignition of flammable or
combustible atmospheres. Disconnect power before servicing.
Failure to remove power and use Daniel approved components may cause serious injury.
3.4.1 Startup for systems using explosion-proof conduit
1. Assemble conduit to the Transmitter Electronics Enclosure. A conduit seal fitting is
required within 18 inches (457 mm) of the enclosure.
2. Check to make certain that all power to field wiring is turned OFF.
WARNING!
HAZARDOUS VOLTAGE INSIDE
Do not open the Transmitter Electronics Enclosure when an explosive gas atmosphere is
present. Disconnect equipment from supply circuit before opening the enclosure.
Failure to remove power may result in serious death or injury.
3. Remove the end cap nearest the conduit entry to gain access to the transmitter
electronics.
4. Pull the wires into the electronics enclosure. Complete the field connection wiring as
shown in see and .
5. Complete the field connection wiring and apply electrical power to the system.
Electrical installation
Installation manual 49

Table of Contents

Other manuals for Emerson Daniel 3410 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson Daniel 3410 Series and is the answer not in the manual?

Emerson Daniel 3410 Series Specifications

General IconGeneral
BrandEmerson
ModelDaniel 3410 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals