4. Remove the transformer retainer using a 1 1/8” wrench and then disconnect and
remove the transformer module (Figure 3-1).
Note
T-21 transducers use W-01 transformers and T-22 transducers use W-02 transformers.
T-21 and T-22 transducer assemblyFigure 3-1:
A. Transducer cable (max. length 15 ft.)
B. Retainer clips
C. Transformer retainer (Standard P/N 1-360-01-958 or High Temperature P/N
1-360-01-978)
D. Transformer module T-21/T-41 (W-01 P/N 1-360-03-090) or T-22 (W-02 P/N
1-360-03-110)
E. Transducer holder
F. Transducer holder O-rings
G. Set screw holes (end of transducer holder)
H. Transducer assembly
I. Mount and backup O-ring
5. Loosen the T-Slot transducer holder assembly with a 1 1/4” socket. Carefully remove
the T-Slot transducer assembly.
Meter repairs
16 Gas Ultrasonic Flow Meter