Configuration and Use Manual 103
Measurement Performance
Measurement Performance TroubleshootingCustody TransferCompensation
Figure 10-2 Meter verification test, original version – Display
Select
OFF-LINE MAINT
Scroll and Select simultaneously
for 4 seconds
Scroll
Select
Scroll
SENSOR VERFY
OUTPUTS
STOP MSMT/YES?
. . . . . . . . . . . . . . . x%
(2)
UNSTABLE FLOW
(1)
Select
PASS ABORTCAUTION
Scroll Scroll
RERUN/YES?
Yes
Scroll
No
Scroll
Select
Correct condition
Correct condition
Select
Scroll
Scroll
SENSOR EXIT
Scroll
Choose output setting
(1) Either Unstable Flow or Unstable Drive Gain may be displayed, indicating that the
standard deviation of the flow or drive gain is outside limits.
(2) Represents the percentage completion of the procedure.