Configuration and Use Manual 139
Troubleshooting
Measurement Performance TroubleshootingCustody TransferCompensation Measurement Performance TroubleshootingCustody TransferCompensation Measurement Performance TroubleshootingCustody TransferCompensation Measurement Performance TroubleshootingCustody TransferCompensation
A027 Security Breach Security Breach Weights and Measures security seal has been
broken. Alarm can be cleared by user, but
authorized procedure is required to reestablish
security. See Chapter 11.
A028 Sensor/Xmtr
Communication Failure
Core Processor Write
Failure
Cycle power to the meter.
The transmitter might need service or upgrading.
Contact Micro Motion. See Section 12.3.
A031
(2)
Undefined Low Power The core processor is not receiving enough power.
Check the power supply to the transmitter, and
check power wiring between the transmitter and the
core processor (4-wire remote installations only).
A032
(3)
Meter Verification Fault
Alarm
Meter
Verification/Outputs In
Fault
Meter verification in progress, with outputs set to
fault. Allow the procedure to complete.
If desired, abort the procedure and restart with
outputs set to last measured value.
A032
(4)
Outputs Fixed during
Meter Verification
Meter Verification In
Progress and Outputs
Fixed
Meter verification in progress, with outputs set to
Fault or Last Measured Value. Allow the procedure
to complete.
If desired, abort the procedure and restart with
outputs set to Continue Measurement.
A033
(2)
Sensor OK / Tubes
Stopped by Process
Sensor OK / Tubes
Stopped by Process
No signal from LPO or RPO, suggesting that sensor
tubes are not vibrating. Verify process. Check for air
in the flow tubes, tubes not filled, foreign material in
tubes, or coating in tubes.
A034
(4)
Meter Verification Failed Meter Verification Failed Rerun the test. If the test fails again, see
Section 10.3.4.
A035
(4)
Meter Verification
Aborted
Meter Verification
Aborted
If desired, read the abort code, see Section 10.3.4,
and perform the appropriate action.
A100 Primary mA Output
Saturated
Primary mA Output
Saturated
See Section 12.18.
A101 Primary mA Output
Fixed
Primary mA Output
Fixed
Check the HART polling address. See
Section 12.19.
Exit mA output trim. See Section 5.4.
Exit mA output loop test. See Section 5.3.
Check to see if the output has been fixed via digital
communication.
A102 Drive Over-Range /
Partially Full Tube
Drive Overrange Excessive drive gain. See Section 12.25.3.
Check the sensor coils. See Section 12.27.
A103 Data Loss Possible Data Loss Possible (Tot
and Inv)
Cycle power to the meter.
View the entire current configuration to determine
what data were lost. Configure any settings with
missing or incorrect data.
The transmitter might need service. Contact Micro
Motion. See Section 12.3.
A104 Calibration-In-Progress Calibration in Progress Allow the meter to complete calibration.
A105 Slug Flow Slug Flow See Section 12.17.
A106 Burst Mode Enabled Burst Mode Enabled No action required.
A107 Power Reset Occurred Power Reset Occurred No action required.
Table 12-5 Status alarms and remedies continued
Alarm
code Communicator ProLink II software Possible remedy