Status alarms and Status Alarm Severity (continued)Table 5-2:
Alarm code Status message Default severity Notes Configurable?
A009 Transmitter Initializing/
Warming Up
Fault
Yes
A010 Calibration Failure
Fault
No
A011 Zero Calibration Failed:
Low
Fault
Yes
A012 Zero Calibration Failed:
High
Fault
Yes
A013 Zero Calibration Failed:
Unstable
Fault
Yes
A014 Transmitter Failure
Fault
No
A016 Sensor RTD Failure
Fault
Yes
A017 T-Series RTD Failure
Fault
Yes
A018 EEPROM Error (Transmit-
ter)
Fault
No
A019 RAM Error (Transmitter)
Fault
No
A020 No Flow Cal Value
Fault
Yes
A021 Incorrect Sensor Type (K1)
Fault
No
A022 Configuration Database
Corrupt (Core Processor)
Fault
Applies only to flowmeters with the
standard core processor.
No
A023 Internal Totals Corrupt
(Core Processor)
Fault
Applies only to flowmeters with the
standard core processor.
No
A024 Program Corrupt (Core
Processor)
Fault
Applies only to flowmeters with the
standard core processor.
No
A025 Boot Sector Fault (Core
Processor)
Fault
Applies only to flowmeters with the
standard core processor.
No
A026 Sensor/Transmitter Com-
munications Failure
Fault
No
A028 Core Processor Write Fail-
ure
Fault
No
A031 Low Power
Fault
Applies only to flowmeters with the
enhanced core processor.
No
A032 Meter Verification in Pro-
gress: Outputs to Fault
Varies Applies only to transmitters with
Smart Meter Verification.
If outputs are set to Last Measured
Value, severity is Info. If outputs are
set to Fault, severity is Fault.
No
A033 Insufficient Right/Left Pick-
off Signal
Fault
Applies only to flowmeters with the
enhanced core processor.
Yes
A034 Meter Verification Failed
Informational
Applies only to transmitters with
Smart Meter Verification.
Yes