Specifications and Reference Data
April 2017
Reference Manual
00809-0100-4021, Rev JA
Specifications and Reference Data
188
Function blocks
Resource block
Contains physical transmitter information including
available memory, manufacture identification,
device type, software tag, and unique identification
Plantweb
Alerts enable the full power of the PW
digital architecture by diagnosing instrumentation
issues, communicating the details, and
recommending a solution
Transducer block
Contains the actual temperature measurement data,
including sensor 1, sensor 2, and terminal (body)
temperature
Includes information about sensor type and
configuration, engineering units, linearization,
range, damping, and diagnostics
LCD display block (when an LCD display is
used)
Configures the local display
Analog input (AI)
Processes the measurement and makes it available
on the F
OUNDATION Fieldbus segment
Allows filtering, engineering unit, and alarm changes
PID block (provides control functionality)
Performs single loop, cascade, or feedforward control in
the field
Block Execution time
Resource N/A
Transducer N/A
LCD display Block N/A
Advanced Diagnostics N/A
Analog Input 1, 2, 3 60 milliseconds
PID 1 and 2 with Autotune 90 milliseconds
Input Selector 65 milliseconds
Signal Characterizer 45 milliseconds
Arithmetic 60 milliseconds
Output Splitter 60 milliseconds

Table of Contents