89
Reference Manual
00809-0100-4021, Rev JA
HART Commissioning
April 2017
HART Commissioning
2. Select pipe material.
3. Select line size.
4. Select pipe schedule.
5. If process Pipe Material, Line Size, or Pipe Schedule is not available under Rosemount X-well Process
selection, select Rosemount X-well Custom sensor type.
6. Enter Pipe Conduction Coefficient. If coefficient is not known, contact factory with pipe material and
pipe wall thickness of application. A custom pipe coefficient will be provided to input into transmitter.
7. Confirm Rosemount X-well Technology Configuration: Fast Keys 2-2-1-11-3