9.1.4  Total power loss
In the event that the transmitter will not power up at all, check the incoming
power supply to make sure power is being delivered to the transmitter.
If the incoming power supply is good, then check fuses F1 and F6 in the
electronics housing. Refer to Figure 9-1 for fuse locations.
Figure 9-1: Fuse Locations
A. F6
Neutral (N)
10 Amp, 250 Vac
B. F1
Line (L1)
10 Amp, 250 Vac
C. F3
O
2
and COe heater
4 Amp, 250 Vac
D. F4
Sample block heater
8 Amp, 250 Vac
Troubleshoot
Manual
December 2022 00809-0500-4880
214 Emerson.com/Rosemount