5. Refer to Figure 10-3 and make sure all test gas lines and electrical
connectors are complete.
6. Install the sensor housing cover.
7. Restore power to the system. Allow the transmitter to reach normal
operating temperature.
8. Insert and bolt the sensor housing in the stack and install insulation.
9. Turn on the test gases at the cylinders and open the instrument air supply
valve.
Remove remote electronics housing
Procedure
1. Turn off power to the system.
2. Shut off the test gases at the cylinders and close the instrument air supply
valve, Figure 10-3.
3. Disconnect the test gas and instrument air lines from the remote
electronics housing.
4. Remove the cover from the electronics housing to expose the electronics
housing terminal blocks, Figure 10-5.
Manual Maintenance and service
00809-0500-4880 December 2022
Rosemount OCX8800A 245