AirFlow
®
S1
Replacement of the water lter cartridge
Unscrew the cap of the water lter (38) on the back side
of the unit.
Unscrew the water lter cartridge using the hexagonal
spanner (10).
Change the water lter spare parts set (39).
Replacement of the complete water lter
Dismantle the uniclamp (92) and disconnect the blue water
hose (70) from the water lter.
Unscrew the external nut of the water lter and put in place
a new complete water lter (38).
38
10
External nut
92
70
Water lter cap
39