multi X 2500 Transport and storage
117
5. Tilt the combustion furnace into vertical position to access
the interfaces in the device interior.
6. Remove the grounding conductor of the combustion furnace
from the floor plate connection.
7. Detach the connectors from the slots.
Connector of the flame sensor (1)
Connector of the combustion furnace (2)
Connector of the thermocouple (3)
To do this, slightly lift the gray lever on the connector of the
furnace (2) (see circle).
8. Tilt the combustion furnace into horizontal position.
9.
If required, remove the auto-protection assembly from the
combustion furnace ( "
Servicing the auto-protection
assembly", p. 96).