multi X 2500 Function and layout of the multi X 2500
27
Fig. 16 Auto-protection assembly
1 Hose 11 (compressed argon gas for pneumatic seal)
2
Hose 8 (argon for safety purging)
Heated gas transfer line to the chlorine module
4 Knob of the clamping mount
5
Valve assembly with integrated filter and
pneumatic seal
Cable for electrical connection
The coupling connects the combustion tube and the measuring gas transfer line. The
combustion tube is inserted into the auto-protection assembly and sealed with a
pneumatic seal. The pneumatic seal is operated with argon. The pneumatic seal is
opened or closed with a toggle switch.
Toggle switch down: Combustion tube sealed
Toggle switch up: Combustion tube released (e.g. for dismantling)
Fig. 17 Toggle switch to lock/release combustion tube
seal